RVS-200 Integrated Modular Wall Diagnostic Station

RVS-200 is a versatile, integrated modular wall diagnostic station which like the RVS-100 measures pulse rate (PR), oxygen saturation (SpO2), non-invasive blood pressure (NIBP) and temperature (oral / axillary probe) with the additional option to add ophthalmoscopes, otoscopes and dermatoscopes. The flexible, modular design offers many configurations and settings to suit the needs of hospitals and clinics as well as office based and long-term care settings.

Integrated RVS-100 Vital Signs Monitor

tick-icon

Option for up to 5 instrument heads

tick-icon

Continuous Monitoring

tick-icon

Spot Check

tick-icon

Triage

tick-icon

Demo Mode

 

Continuous Monitoring is designed for monitoring patients over time and includes physiological and technical alarms. Spot Check is designed for taking a single set of vital signs measurements on a patient. Patient information can be entered and managed, and while technical alarms are still available, physiological alarms are disabled. Triage is designed for rapidly taking vital signs measurements on many patients. Patient information is disabled, in addition to physiological alarms. Demo mode demonstrates how each measurement mode works.1

Image gallery

Riester RVS-200 web

RVS-200 with NIBP blood pressure, SpO2 pulse oximeter and oral probe thermometer (1961-RRBXE).

Hardware

Features and benefits – hardware

  • Large 8“ full HD colour TFT-LCD screen displays large numbers and bright colour-coded parameters so that vital signs can be seen clearly and interpreted easily.

  • Screen is visible in both dark and lit rooms.

  • The touch screen and intuitive user interface makes it quick and easy for patient information to be entered and settings amended as required.

  • The modular design is chosen by the customer around their needs, starting with the choice between a table/mobile stand or wall mounted, through to the optional accessories like a printer.

  • Can be plugged into the mains or powered off the battery for 17 hours.2

  • Riester is committed to providing reliable clinical measurements, including the choice of SpO2:

    • Riester
    • Nellcor
  • Centralised power supply supports up to 3 diagnostic handles, from a choice of Riester ri-scope® L and EliteVue instrument heads:

  • Integrated specula dispenser for the otoscope and nasal specula with 5 tubes for 5 different sizes of specula, dispenses a new specula for each examination and is easy to re-fill and disinfect.

Software

Features and benefits – software

  • RVS-100 is programmed with EWS: NEWS, MEWS and 5 customisable profiles which can be set by the end-user if they have bespoke requirements.
  • Serve a large patient population with the ability to save up to 1000 patient records, with up to 5000 measurements per patient.
  • Can be used on all patient types: adult, paediatric and neonate.
  • RVS-100 can communicate through a wired or wireless connection to the hospital’s EMR/EHR, according to the HL7 standard.3
  • Software languages available: Chinese, Danish, English, French, German, Italian, Polish, Portuguese, Spanish (Europe) and Traditional Chinese

Product details

 

Space requirement
RVS-200  
Space requirement

600mm = 23,622 inch, 309mm = 12,165 inch, 871mm = 34,291 inch, 100mm = 3,937 inch

Power supply 100 - 240 V AC, 50/60 Hz
Battery Lithium-Ion battery, rechargeable, 10.8 V DC, 6600 mAh. Charging time to a 100%: 6 hours, Operating time approx. 11 hours (new and fully charged battery at 25 °C ambient temperature with connected SpO2, Temp, and NIBP on AUTO mode (15 minutes interval), Low battery indicator, Charging battery indicator
Monitor dimensions 303 x 230 x 110 mm (11.93 x 9.05 x 4.33 in)
Monitor weight 3600 g (7.94 lb) inc. temperature module, printer
Basket dimensions 270 x 225 x 200 mm (10.63 x 8.86 x 7.87 in)
Alarm Visible LEDs (Yellow, Red, Blue, Green, Orange) Audible Speaker, Gives audible alarm, QRS tone. Supports Pitch Tone and multi-level volume. Alarm tones meet the requirement of IEC 60601-1-8. Alarm Pressure: 45 dB to 85 dB test distance is 1 meter from the tone
Ports 1 x AC power inlet
1 x standard RJ45 interfaces.100 BASE-TX, IEEE 802.3
4 x USB ports
1 x Equipotential grounding point
1 x RJ11 connector for nurse call
DC out port 15 V/1.2 A

 

NIBP (non-invasive blood pressure)
Measuring method Automatic oscillometric
Measurement range Adult SYS 30 - 270 mmHg, DIA 10 - 220 mmHg, MAP 20 - 235 mmHg
Paediatric SYS 30 - 235 mmHg, DIA 10 - 220 mmHg, MAP 20 - 225 mmHg
Neonate SYS 30 - 135 mmHg, DIA 10 - 110 mmHg, MAP 20 - 125 mmHg
Cuff pressure range 0 - 280 mmHg
Resolution 1 mmHg
Pressure accuracy Static ±3 mmHg
Clinic ±5 mmHg, standard deviation ≤ 8 mmHg
Pulse rate range 40 - 240 bpm
Pulse rate accuracy 1 bpm
Initial inflation pressure Adult 160 mmHg, Paediatric 130 mmHg, Neonate 75 mmHg
Double overpressure protection (hardware and software).
Adult 297 ±3 mmHg, Paediatric 252 ±3 mmHg, Neonate 147 ±3 mmHg
Alarm range SYS 0 - 300 mmHg
DIA 0 - 300 mmHg
MAP 0 - 300 mmHg

 

SpO2 – Riester  
Measurement range 0 - 100%
Resolution 1%
Accuracy ±2% (70% to 100%)
±3% (40% to 69%)
Unspecified (0% to 39%)
Alarm range 0 - 100%
Average time 4 sec, 8 sec, 16 sec
Pulse rate
Measurement range
20 - 250 bpm
Pulse rate Resolution 1 bpm
Pulse rate Accuracy ±1% or ±1 bpm which ever is greater

 

SpO2 – Nellcor  
Measurement range 0 - 100%
Resolution 1%
Accuracy 70 - 100%: ±2% (adult, paediatric)
70 - 100%: ±3% (neonate)
0 - 69%: unspecified
Alarm range 0 - 100%
Average time 8 sec, 16 sec
Pulse rate
Measurement range
20 - 300 bpm
Pulse rate Resolution 1 bpm
Pulse rate Accuracy 20 - 250 bpm: ±3 bpm
251 - 300 bpm: unspecified

 

SpO2 – Masimo  
Measurement range 0 - 100%
Resolution 1%
Accuracy 70 - 100%: ±2% (adult, paediatric, non-motion conditions)
70 - 100%: ±3% (neonate, non-motion conditions)
70 - 100%: ±3% (motion conditions)
0 - 69%: unspecified
Alarm range 0 - 100%
Average time 2 - 4 sec, 4 - 6 sec, 8 sec, 10 sec, 12 sec, 14 sec, 16 sec
Pulse rate
Measurement range
25 - 240 bpm
Pulse rate Resolution 1 bpm
Pulse rate Accuracy ±3 bpm (non-motion conditions)
±5 bpm (motion conditions)

 

RPT-100 (Filac 3000) Predictive Thermometer
Oral (Quick Mode) non-fever temps: 3 - 5 seconds, fever temps: 8 - 10 seconds
Oral (Standard Mode) 6 - 10 seconds
Axillary Mode 8 - 12 seconds
Rectal Mode 10 - 14 seconds
Direct Mode (all sites) 60 - 120 seconds
Measurement accuracy 35.5 °C - 42 °C
Direct Mode (all sites): ±0.1 °C (±0.2 °F)
Standard Mode (all sites): ±0.1 °C (±0.2 °F)
Quick Mode (oral only): ±0.3 °C (±0.5 °F)

Ordering information

RVS-200

RVS-200 includes an extension module with 2 handles, adult cuff and adult SpO2 sensor. Instrument heads must be ordered separately. Further handles may be added using extension modules.

ART. NO. REGION DESCRIPTION
1961-RRXXE EU RVS-200, NIBP (Riester), SpO2 (Riester), EU plug
1961-RRBXE EU RVS-200, NIBP (Riester), SpO2 (Riester), Pred. Temp, EU plug
1961-RRBPE EU RVS-200, NIBP (Riester), SpO2 (Riester), Pred. Temp, Printer, EU plug
1961-RNXXE EU RVS-200, NIBP (Riester), SpO2 (Nellcor), EU plug)
1961-RNBXE EU RVS-200, NIBP (Riester), SpO2 (Nellcor), Pred. Temp, EU plug
1961-RNBPE EU RVS-200, NIBP (Riester), SpO2 (Nellcor), Pred. Temp, Printer, EU plug
1961-RRXXU US RVS-200, NIBP (Riester), SpO2 (Riester), US plug
1961-RRBPU US RVS-200, NIBP (Riester), SpO2 (Riester), Pred. Temp, Printer, US plug
1961-RNXXU US RVS-200, NIBP (Riester), SpO2 (Nellcor), US plug
1961-RNBXU US RVS-200, NIBP (Riester), SpO2 (Nellcor), Pred. Temp, US plug
1961-RNBPU US RVS-200, NIBP (Riester), SpO2 (Nellcor), Pred. Temp, Printer, US plug

These configurations are standard. Further configurations are available for country/regional specifications, please contact your Riester distributor for further information.

Accessories

ART. NO. REGION DESCRIPTION
1840-BT All Non-contact thermometer ri-thermo® sensioPRO+ Bluetooth
1835-BT All Tympanic thermometer ri-thermo® tymPRO+ Bluetooth
12634 All except US Probe cover for ri-thermo® tymPRO 1 rack/10 pcs.
12640 All Probe cover for ri-thermo® tymPRO 30 racks/300 pcs.
13315 All Barcode reader
13316 All Wi-Fi for RVS-100
13317 All Mobile stand with basket
13324 All except US Wall mount basket
13333 All except US NIBP extension tube with connector
13321 All except US Paper roll for printer RVS-100, pack of 20 pcs.
12668 All Rectal Probe Filac 3000, length 2.7 m
12669 All Oral / Axillary Probe Filac 3000, length 2.7 m
12688 All Single probe covers predictive thermometer, pack of 500 pcs.
13347 All Bluetooth USB-Dongle for RVS-100

Cuffs

ART. NO. REGION DESCRIPTION
107.005 All Velcro cuff (adult) 54.5 x 14.5 cm, 1 tube, black
122.005 All Velcro cuff (large adult) 70 x 15 cm, 1 tube, black
129.006 All Velcro cuff (children) 35.5 x 10 cm, 1 tube, blue

134.02

All Velcro cuff (newborn) 15.5 x 5 cm, 1 tube, blue

SpO2 Sensors

ART. NO. REGION DESCRIPTION
13300 All SpO2 sensor Riester neonatal for RVS-100
13301 All SpO2 sensor Riester children for RVS-100
13302 All SpO2 sensor Riester adult for RVS-100
13303 All SpO2 sensor Nellcor neonatal for RVS-100
13304 All SpO2 sensor Nellcor children for RVS-100
13305 All SpO2 sensor Nellcor adults for RVS-100

Instrument Heads

ART. NO. REGION DESCRIPTION
10512-301 All EliteVue head only, LED 3.5 V with anti-theft security
10565-301 All ri-scope® L2 otoscope head, LED 3.5 V with anti-theft security
10567-301 All ri-scope® L3 otoscope head, LED 3.5 V with anti-theft security
10594-301 All except US ri-scope® L1 ophthalmoscope head, LED 3.5 V with anti-theft security
10595-301 All ri-scope® L2 ophthalmoscope head, LED 3.5 V with anti-theft security
10596-301 All ri-scope® L3 ophthalmoscope head, LED 3.5 V with anti-theft security

 *Additional instrument heads are available. All ri-scope® L 3.5 V instrument heads are compatible with the RVS-200.

Downloads

rvs-100-productteaser-500x500-crop-1
RVS-100 Vital Signs Monitor

RVS-100 Vital Signs Monitor with versatile, intuitive touch screen with Early Warnings Scores (EWS) and Bluetooth connectivity.

Riester ri-scope L otoscope 05 web
ri-scope® L Otoscope

Riester’s ri-scope® L line of otoscopes with the choice between LED or Xenon illumination.

riester-tympro-c-1-500x500-crop-1-1
ri-thermo® tymPRO+ tympanic thermometer

Tympanic thermometer which measures temperature with single-handed operation and single-use probe covers.

ri-thermo-sensiopro-square-500x500-crop-1-1
ri-thermo® sensioPRO+ thermometer

A thermometer which measures temperature without contact with single-handed operation.

Where to buy?

Riester products are available in over 150 countries. Enter your country to find a Riester distribution partner near you.

number1

Please note: Unit must be restarted to exit Demo mode.

number2

11 hours in continuous monitoring mode.

number3

HL7 (Health Level Seven) is a standard for exchanging health information between medical applications. This standard defines a format for the transmission of health-related information.

number4

At a distance of 15 mm (0.59 in) between landmark and distal tip of the specula.